TOC 2500 OFFLINE ANALYSER

Features

  • Automatic sample introduction with the enter of trace sample, we can control the value of the trace sample. No contamination will during testing which will be harmful to human beings and the environment.
  • It adopts UV Catalytic oxidation during which there is no need to add acid, gas or catalytic with low loss of routine maintenance cost.
  • 7 inches touch screen with an operation panel, which is easy to operate and maintain.
  • The system is set with a password for protection which is in compliance with FDA-21    CFR Part11 Requirements. Also, it is in compliance with USP32-643 requirements.
  • Auto sampler is optional which can be used for testing various different samples without person.
  • Special designed for testing water less than 1.5 ppm. Online and office testing functions can be easily switched.
  • 8GB mass storage memory which does not restrict date and time.
  • Less than 3 minutes for each analysis.
  • Clear historical records which are traceable according to analysis date.
  • Data for a specific whole day can be derived and saved to USB directly.
  • Equipped with a Bluetooth printer which is easy to operate and print quickly.
  • Modular design ensures quick installation and easy maintenance.
  • Optional online monitoring devices meet different requirements.    

 

Specifications

  

 detection range

0.001-1.5mg/l 

 detection limit

1ppb 

 max tolerance

±5% 

 analysis time

3 min 

 response time

within 10 min 

sample temperature 

1-95? 

 environment temperature

10-40? with temperature change ±5 ?/d

 sample flow speed

3ml/min 

 relative humidity

≤85% 

 drifting

±5%

repeatability tolerance

≤3%

drifting  

 ±5%

power 

 220V±10%

 frequency

50HZ+1HZ 

 power

100W 

 dimension

440*220*300mm 

Contact Us Call Us

Share: