Online Fluorescence Dissolved Oxygen Sensor BH-DY06
Online Fluorescence Dissolved Oxygen Sensor BH-DY06 is coated with a luminescent material. Green light from an LED is transmitted to the sensor surface. The green light excites the luminescent material. As the material relaxes it emits red light.
The time it takes for the red light to be emitted is measured. Between the flashes of blue light, a red LED is flashed on the sensor and used as an internal reference. Increased oxygen in the sample decreases the time it takes for the red light to be emitted. The time measurements correlate to the oxygen concentration.
Features of Online Fluorescence Dissolved Oxygen Sensor BH-DY06
- Green excitation fluorescence life detection technology, no electrolyte, no interference, no frequent calibration, stable and reliable measurement.
- Built-in temperature sensor automatic compensation function for water temperature.
- Support MODBUS/RS485 & 4-20mA current interface output.
Model | Parameter |
Measuring Range | 0~20 mg/L or 0~200% saturation |
Accuracy | ±5% or ±0.3mg/L |
Response Time | 90s |
Temperature Sensor | DS18B20 digital temperature sensor |
Working Temperature | 5~40℃ |
Power Supply | 12VDC±20% |
Calibration | One point or two point calibration |
Protection Grade | IP68 |
Size | Φ34×222mm |
Analog Interface | 4-20mA (optional) |
Pressure Of Work | <5bar |
Digital Interface | MODBUS /RS485 |
Cable Length | 8m |
Lifetime Of Fluorescent Cap | 1 year |
Material | The main body is stainless steel 316 and the sensitive film is silicone |