Optical Turbidity Sensor
- Product Name: Optical Turbidity Sensor V2.0
- Product brand:FLUIDINGS
Fluidings measures turbidity with an fiber optics sensor. An 860nm light from the emitter enters the sample via fiber and scatters off particles in the water. The scattered light at 90 degrees, enters a PD
detector via fiber. This follows the nephelometric technique of measurement, and values are expressed in nephelometric turbidity units (NTUs). It follows the ISO7027 standard.
Features of Optical Turbidity Sensor
- Ambient light rejection technology
- Internal temperature compensation for optics
- Easy to calibrate
- Low calibration solution volume, 50mL minimum
- Provide SS and transparency output(calculated from turbidity)
- Support inside leakage alarm
- RS485 Modbus RTU output
Measurement Method | scattered light at 90 degrees, ISO7027 |
Range | 0.3~100NTU, 0.3~1000NTU, 1.0~4000NTU |
Resolution | 0.01NTU, 0.1NTU, 1.0NTU |
Accuracy | ± 5% reading or 1 NTU w.i.g |
Operating temperature | 0~45℃ |
Storage Temperature | -10~50℃ |
Min. Detection Limit | 0.3NTU |
Warranty | 1 year |
Depth | IP68, 10m Max |
Power | DC 5~12V 10mA (normal) |
Output | RS485 and Modbus protocol |
Materials | PVC, PMMA, SS316(Default), TC4 is optional |
Dimensions | Length 187mm, diameter 22mm |
Flow rate | < 3 m/s |
Response time | Minimum 10s T90 |
Field life* | Sensor 2 years or greater |
Recommended Calibration maintain Frequency* | Sensor 6 months |